# What you are
checking:
Procedure Frequency What to record Troubleshooting
T1 Transducer
operation
Check that all 4 transducers are
humming when running
Daily Write down the check
date
T2 Transducer
lifecycle (200
hours)
Run Check on Manual mode 10% -
100% output. At 100% the hand
set should run almost silent. Let
the handset run for 2-3 minutes.
Monthly Write down the check
date & any observation
If the transducers are noisy, it’s
time to replace as the
transducers are showing signs
of reduced output.
If a handset is running warmer
than usual this is also a sign
that it might be time to replace.
If there is visible or suspected
damage to the cavitation wire
turn off and disconnect from
the machine immediately and
replace hand set.